In case we have not met, I’m Alexa Williams, Ask Alexa. I have worked here at Etelux for almost 15 years establishing great relationships with our customers and partners. Throughout my career at Etelux, I have had the pleasure of advocating for you, the customer, to make sure that you are taken care of. Most of the time, I am the friendly voice that answers the corporate phone line, so if we have not met in person, I’m sure we have talked. This form is to reach me directly, and I will respond as soon as I can. You can ask me anything, such as questions about our enclosures, help with which replacement parts to order, certification questions, etc…
Etelux is a global leader in the manufacture and design of glove box technology products.
Serving the Chinese scientific industry for over 15 years, discover how quality products, competitive pricing, and next day delivery allow us to achieve our mission “Enabling Technology for a Better World ".
Ultricies eu consectetur vitae donec scelerisque ante a ut morbi aptent tincidunt est non sagittis ut.
The vacuum pump shall be kept open during system operation.
When the oxygen concentration in the tank exceeds 200 ppm, the equipment maintenance personnel shall be informed to operate the system, otherwise the water and oxygen purification system will be damaged.
Before starting the cycle, check whether the box condition or regeneration is in progress. If the above mentioned operation is in progress, it is absolutely forbidden to cycle.
two purification columns cannot be regenerated at the same time.
Before regeneration, ensure that the vacuum glove box has been cleaned by inert gas (H2O < 200ppm) to ensure sufficient regeneration gas supply. The pressure of regeneration gas is 0.06-0.08MPa.
Before starting regeneration, please confirm that the corresponding purification column cycle is closed.
Regeneration process needs 24 hours. Once the execution cannot be disconnected, 3-6 hours after the start of regeneration, use regeneration gas to pay attention to the amount of regeneration gas, adjust the pressure of regeneration gas properly, and ensure the sufficient supply of regeneration gas in these 3 hours. After 6 hours, the regeneration origin can be closed.
During the regeneration process, the vacuum pump cannot be turned off. Do not touch the top of the purification column to prevent scalding.
Once the regeneration is terminated unexpectedly, please ensure that the regeneration and circulation of the purification column cannot be restarted within 12 hours, otherwise the equipment will be damaged.
During the operation, pay attention to protect the operation gloves from damage.