Open the shipping box, first check the combination of the vacuum glove box in the transportation process. If there is loose, tighten the corresponding screws, then you can try the empty.

First, vacuum and inflatable ingress:

1, this product is divided into two parts: the main chamber and the antechamber. The main chamber cannot be empty alone, and the antechamber can be empty alone.

2, first open the door inside the transition room, then close all gloves interface cover, valve and transition outdoor door.

3, turn on the three-way valve (counterclockwise rotation) between the gloves, cover the space between the glove interface and the gloves, so that inside and outside the glove (chamber and interface) at the same time, keep it both air pressure balance.

4, connect the vacuum pump to a valve on the antechamber, open the vacuum pump, slowly open this valve, pumping the system (if the speed of opening the valve is too fast, may cause gloves expand).

5, when the vacuum table pointer drops and stabilizes in -0.1MPa, the vacuum is complete, and the valve should be turned off first.

6, then charge the inert gas into the chamber through another valve until the air pressure is slightly higher than the atmospheric pressure (the air pressure gauge index is 0 MPa), so that the pressure in the chamber, the pressure is substantially balanced, and the internal is slightly higher than the outside, which can guarantee the atmosphere not penetrate the interior of the chamber (if you want to make the gas in the chamber, it is best to repeat the above pumping and inflatable process several times). Turn off the three-way valve connecting the glove interface, open the cover on the glove interface, you can operate.

Second, the object is placed in and remove the main chamber: 1, the object is placed in the main chamber: When you complete the above “(1) Different air and inflatable, if you need to put the object to the chamber, first put the interior of the antechamber, open the outer door of the antechamber, put the object into the outside, then close the door, and use the two valves on the antechamber to pumping and inflatable. Be sure to balance the pressure in the main chamber and antechamber and the atmospheric pressure can open the door inside and take the object into the main chamber. 2. Methods taken out in the body chamber: first put the object into the antechamber, and close the antechamber indoor. If the air pressure and atmospheric pressure in the antechamber are basically balanced, the antechamber outdoor can be removed.