Quick Summary (read this first)

  • Stabilize pressure (target +3–8 mbar) and protect people first.
  • Identify the trigger: O₂ spike, dew-point spike, both, spill, tear, or power loss.
  • Log time zero, then follow the scenario SOP below.
  • Acceptance to return to service: O₂ back to ≤1–5 ppm, dew point ≤ −60/−40 °C (as your spec) and stable trends for 2–4 h.

1) O₂ Spike (dew point normal) — Likely air ingress

Typical causes: bad antechamber equalization, door cracked open, glove yanked.

Do now

  1. Keep circulation ON; hold +5 mbar.
  2. Audit last transfer: was the load-lock equalized (±1 mbar) before inner door?
  3. Run a short purge (box gas) if O₂ > setpoint for >10 min.
  4. Leak check fast: glove beads, door O-rings, feedthrough caps.

Then

  • If O₂ decays normally, no regen yet.
  • If plateau persists, schedule purifier regeneration (≤5% H₂/N₂) after confirming exhaust and interlocks.

Accept back to service: O₂ ≤ target and falling/stable for 2–4 h.


2) Dew-Point Spike (O₂ flat) — Solvent vapor or wet load

Typical causes: open beakers, wet wipes left inside, hot parts off-gassing.

Do now

  1. Close lids; bag solvent-soaked wipes; remove unnecessary sources.
  2. Cold/charcoal traps online and at temperature; check trap seals.
  3. Maintain circulation; avoid high purge that spreads vapor.

Then

  • If DP remains high, run a moderate purge and consider regen only after traps are working (don’t waste purifier media).
  • Review tool placement: keep solvent work downstream of sensor intake.

Accept back to service: dew point ≤ target and stable for 2–4 h.


3) Both O₂ and Dew Point Spike — Door fully open or major breach

Do now

  1. Secure doors, set +5 mbar, circulation ON.
  2. Short purge in increments; watch trends (avoid relief hissing).
  3. Inspect gloves, windows, large feedthroughs; replace/retorque as needed.

Then

  • If recovery is slow vs historical, regenerate (after verifying exhaust/interlocks).
  • Run a pressure-decay leak test once numbers settle.

Accept back: O₂ & DP within spec and flat for 2–4 h; leak rate ≤ your limit (e.g., ≤1×10⁻² mbar·L/s for general R&D).


4) Glove Tear or Port Damage

Do now

  1. Clamp/cover the port immediately (cap ring, blank plate, or taped film).
  2. Lower setpoint to +3–5 mbar to reduce flow through the defect.
  3. Replace the glove: clean bead & O-ring, re-seat evenly; quick leak check.

Then

  • Trend O₂/DP for 30–60 min; run a short purge if needed.
  • Record part number, thickness, operator for traceability.

Accept back: stable O₂/DP; no pressure creep; glove passes quick pinch test.


5) Solvent Spill Inside the Box

Do now

  1. No wiping frenzy. Cover with inert absorbent (pads, dry spill granules).
  2. Cold/charcoal traps ON; maintain +5 mbar.
  3. Remove saturated absorbent to a sealed can; minimize turbulence.

Then

  • Moderate purge in steps if DP/odor remains high.
  • Consider regen only after traps and cleanup are complete.

Accept back: DP returns to baseline; odor absent; traps serviced/logged.


6) Power Loss / Controller Crash

Do now

  1. Verify pressure relief path; prevent over/under-pressure on restart.
  2. After power returns: bring up circulation first, then sensors, then optional purge.
  3. Validate sensors (O₂ zero or cross-check; DP dry check) before declaring normal.

Then

  • Inspect logs for gaps; annotate event; check UPS coverage for controllers/sensors.

Accept back: control stable; sensors sane; O₂/DP trending to baseline.


Minimal Decision Table

SymptomLikely causeFirst actionSecondary
O₂↑, DP normalAir ingressCirculation ON, check load-lock SOPShort purge → regen if plateau
DP↑, O₂ normalSolvent vaporLids on, traps ONModerate purge → adjust workflow
O₂↑ & DP↑Major door/glove issueSecure doors, +5 mbar, short purgeLeak test → regen
Rapid pressure swingsPID/relief/valveSet +5 mbar, soften PIDClean/replace valve filter
Odor persistsTrap saturatedSwap charcoal/cold trapRegen purifier
Readings erraticSensor/bypassVerify sampling in main loopZero/span or cross-check

Acceptance Targets (edit to your site)

  • Pressure: +3–8 mbar steady (variance ≤1 mbar at idle).
  • O₂:1–5 ppm and flat/down for 2–4 h.
  • Dew point:−60/−40 °C (per process) and stable.
  • Leak rate:1×10⁻² mbar·L/s (or your spec).

Post-Incident Prevention (what to change tomorrow)

  • Antechamber discipline: three-cycle evac to ≤10 mbar abs, equalize ±1 mbar before inner door.
  • Traps: place near sources + in the circulation return; set service cadence.
  • Glove care: daily pinch test; keep spares at station; inspect beads & O-rings weekly.
  • Logging: 7-day trends for O₂/DP/pressure; note time-to-recover after transfers.
  • Training: one laminated 90-second playbook attached to the box.